HC2-LDP

Liên hệ

Field of Application
The low dew point probe can be used in compressed air systems, refrigeration driers and adsorption driers. ROTRONIC offers the probe with EX certificate for environments with ATEX requirements.

Accessories
Suitable measurement chambers with a defined air flow rate and a quick connector for optimum dew point measurement are available in POM and stainless steel. The base unit can be used for customization.

  • Highly accurate dew point and temperature measurement
  • Withstands pressures up to 100 bar
  • HYGROMER® LDP-1 sensor element
  • Excellent repeatability
  • Trace moisture monitoring at low limits
  • With the latest AirChip4000 measurement technology
  • Compatible with HF5 / HF8 / PF4 transmitters and HP22 / HP23 handhelds
Probe typeDigital dew point & temperature probe
Dimensions145 x 140 mm (probe head)
Hexagonal nut: 3 x 140 mm
Weight260 g
Measuring range-70…85 °C Td / -40…85 °C
Dew point accuracy @ 23 °C±2 K Td (-50…20 °C Td)
±3.5 K Td (-60…-50 °C Td)
Temperature accuracy @ 23 °C±0.2 K
Dew point sensorHYGROMER® LDP-1
Long-term stability dew point sensor<1 °C Td/year
Response time dew point sensor Ƭ63: @ -50 °C TdTypically <10 min, with filter
Response time temperature sensor Ƭ63: @ -50 °C TdTypically <15 min, with filter
Pressure resistance16 bar
100 bar (with LDP-MCS measurement chamber)
Power supply / Power consumption3.3…5 VDC / <1.5 mA
Temperature sensorPt1000, Class B 1/3
Filter cartridgeSintered stainless steel, 50 μm
Measurement interval2 seconds
Communication interfacesUART
HousingStainless steel 1.4301
IP protection ratingIP65
FDA/GAMP compatibilityYes
CE / EMC conform2004/108/EC
MaintenanceAnnual calibration (recommended)
Storage and transport conditions-40…85 °C, non-condensing
 
Nhà cung cấp

Chưa cập nhật

Xuất xứ

Chưa cập nhật

Đánh giá

Chưa có đánh giá nào.

Hãy là người đầu tiên nhận xét “HC2-LDP”